Customization: | Available |
---|---|
Classification: | Physiological Functions of Diagnosis and Monitoring Equipment |
Type: | Digital Imaging System |
Still deciding? Get samples of US$ 1850/Piece
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
l Automatic storage of test results, review and print all parameters in test results;
l Color touch screen, all parameters and histograms can be displayed in the same screen;
l Support the expansion of external devices such as mouse, keyboard and printer;
l Optical five reference ranges: newborn, children, adult women, adult men, and general use;
l Multiple clogging elimination methods: high-pressure burning jewel hole, high-pressure reverse rinsing and soaking;
l Automatic cleaning of sample probe;
l By VGA interface, external projector or external display screen can be connected;
l Chinese and English system language;
l Built-in thermal printer, serial port printing and other printing methods;
l With quality control management and calibration management function;
l Use the cyanide-free SFT method (environmentally friendly) to measure hemoglobin;
l Floating boundary algorithm and histogram abnormal prompt function.
l Test parameters: WBC 3-part differential counting, 23 parameters (including WBC, RBC, PLT color histogram);
l Measurement principle: counting by electrical impedance method, colorimetric method to measure HGB;
l Quality control method: L-J, automatic drawing and printing of quality control charts;
l Sample volume: trace detection, 10μL peripheral blood or anticoagulant blood, pre-dilution mode 20μL;
l Precision: WBC (white blood cell) CV≤4.0%, RBC (red blood cell) CV≤2.0%, HGB (hemoglobin) CV≤2.0%, PLT (platelet) CV≤8.0%, MCV (average red blood cell volume) CV≤3.0% ;
l Accuracy: the allowable relative deviation range: WBC≤±15%, RBC≤±6.0%, HGB≤±6.0%, PLT≤±20.0%, HCT (hematocrit)≤±9.0%;
l Blank count: WBC≤0.5×109/L, RBC≤0.05×1012/L, HGB≤2.0g/L, PLT≤10.0×109/L;
l Carry-over: WBC≤3.5%, RBC≤2.0%, HGB≤2.0%, PLT≤5.0%;
l Linearity deviation: WBC≤±5%, RBC≤±5%, HGB≤±3%, PLT≤±10%;
l Related coefficient: WBC≥0.990, RBC≥0.990, HGB≥0.990, PLT≥0.990;
l Display: color LCD touch screen;
l Detection channel: dual channel;
l Test speed: 35 (or 60) specimens/hour, continuous working for 24 hours;
l Data storage: It can automatically store more than 30,000 groups of complete results (each result has three histograms);
l Interface: RS232 interface, VGA interface;
l Power supply: ~100V-240V; 50/60Hz.
l Mouse
l Thermal print paper
l Power cord
l Grounding cable
l Serial cable
l Pipe assembly (for Diluent, Hemolytic agent,Waste liquid respectively)
l Touch pen
l User manual
l Key of device side board
Physical characteristic
Dimension: 433mm(L)×300mm(W)×430mm(H)
Weight: 19kg
Operating Environment:
Temperature: 18ºC~25ºC
Relative humidity: ≤80%
Atmospheric pressure: 86.0kPa~106.0kPa
Storage Environment:
Temperature: -10ºC~55ºC
Relative humidity: ≤93%
Atmospheric pressure: 50.0kPa~106.0kPa